CO2 incubators are widely used in scientific research to grow and maintain cell cultures.
Typical fields of application include tissue biotechnology, in vitro fertilization, neuroscience, cancer research and other applications in breast cancer research.
Elegant, reliable and intuitive, Esco CelCulture® CO2 Incubators provide complete sample protection and make your scientific dreams come true.
CelCulture® 240L Incubator, IR Sensor, CO2 Control, ULPA, Moist Heat Decon, 230 VAC, 50/60 Hz
Electrogalvanized steel construction with baked epoxy-polyester powder antimicrobial coating paint.
The external surfaces are coated with to eliminate 99.9% of surface bacteria within 24 hours of exposure.
Wet heat decontamination cycle at 90 ° C
Validated by HPA (UK - Health Protection Agency)
ULPA FILTER
• 99.999% efficacy, superior to conventional HEPA filters
• Air filters continuously present
• The chamber returns to ISO Class 5 purity in 11 minutes after closing the door to avoid contamination
SHELVES
• Perforated shelves to improve uniformity (No. 4 standard pieces)
• Anti-overturn
• Stainless steel
• Integrated handle
• Removable without tools for easy cleaning
DIRECT HEAT AND AIR CAVITIES
• Rapid and uniform heating
• Quick temperature recovery
• The cavity improves the stability of the chamber
AIR DUCT
• Directs the airflow for fast recovery and excellent uniformity
• Easily removable for cleaning
WATER TRAY
• Properly heated by a base heater to provide high humidity
• Gentle air flow over the water surface accelerates moisture recovery
ROUNDED CORNERS
• Seamless design
• They facilitate cleaning
SWITCH ON THE DOOR
Automatically turns off the fan and gas supply when the door is opened.
CO2 SENSOR
• Single beam, dual wavelength IR sensor is dispersion free
• Auto-zeroing
SMARTSENSE TM MICROCONTROLLER INTERFACE
Intuitive, in complete control of the monitoring system.
DIGITAL CONTROLLER
Complete, the alarms are configurable by the operator
Temperature
CO2
Humidity and O2 - if installed
CELALERT
The alarm system reminds the operator to replace components.
The intelligent data and event recorder stores all parameters of the incubator with on-screen recall.
The built-in 2 MB memory ensures long-term data retention
TOP COVER
Provides quick access to electrical panel components
Easily adjustable
EXTERNAL DOOR
• Invertible
• Heated to prevent condensation
SAMPLING PORT
It allows to obtain a direct measurement of the chamber atmosphere and also of the temperature and CO2 concentration
INTERNAL GLASS DOOR
To observe the cells in the samples inside the chamber during operation
DOOR LOCK
To lock and unlock the glass door
LEVELING FEET
Easily adjustable
For IVF applications, Esco CelCulture® CO2 incubators are certified as safe for embryos.
Undergoing rigorous technical testing with the Mouse Embryo Assay (MEA), CelCulture® has extraordinarily 100% embryo survival. The Mouse Embryo Assay (MEA) is the de facto standard test performed to demonstrate that a procedure or instrument is safe to use for handling human embryos (e.g. for in vitro fertilization or IVF).
The Esco CelCulture® CO2 Incubator is indicated by the UL Subscriber Laboratory in accordance with the requirements of both the United States and Canadian standards for electrical and mechanical integrity.
Wet heat decontamination cycle at 90 ° C Validated by HPA
The Esco CelCulture® CO2 incubator decontamination cycle has been evaluated as an effective method for the inactivation of fungi, bacterial spores of Aspergillus brasiliensis and Bacillus atrophaeus and vegetative cells of Pseudomonas aeruginosa, Staphylococcus aureus, Staphylococcus epidermidis, Entercherobichia coli and Escherobacteria coli , which are normally hardy.
External dimensions of the product
Height: |
900 |
mm |
Length: |
750 |
mm |
Depth: |
770 |
mm |
Internal dimensions of the product
Height: |
633 |
mm |
Length: |
595 |
mm |
Depth: |
640 |
mm |
Volume : 240L
Shelves
• No. of perforated shelves 4
Temperature
• Temperature control range: from ambient T +7 to 60 ° C
• Temperature uniformity < + 0.5 ° C
• Temperature accuracy < + 0.1 ° C
• Temperature reset (after 30 sec with door open): < 6 minutes
Humidity
• The humidity range exceeds 95% when using the water pan
Infrared CO2 sensor
• CO2 control range: 0-20%
• Accuracy of% CO2: + 0.1%
• Humidity recovery time (after 30 seconds with the door open): < 6 minutes
Diet
• Mains: 220-240 VAC, 50/60 Hz
• Nominal power at 37 ° C: 42.2 W
• Maximum power consumption: 1270 W