The BIOHAZARD ATLANTIC HOOD is a microbiological safety cabinet, with vertical laminar flow of class II type A2, designed for the protection of the operator, the product and the environment, equipped with:
control and programming panel with a 5.7 "TFT LCD touch screen microprocessor that allows the automatic maintenance of the constant speed of the vertical laminar flow and of the suction-front barrier of the air,
front screen in 5 mm thick tempered glass with motorized movement,
2 absolute filters Hepa class H14 (for vertical laminar flow and for expelled air),
fluorescent lighting lamp,
air / vacuum tap,
gas tap with safety solenoid valve,
2 internal hood service sockets, DOP test hose connection,
worktop in stainless steel AISI 304 2B optionally whole, basin or perforated
Accessories on request:
- A2790043 - 30 W germicidal UV lamp (consigliato)
- Activated carbon cell to be cascaded after the Hepa filter for expelled air for processing with toxic vapors
- Perforated work surface, divided into removable and autoclavable sectors
- Ø150 mm external air discharge conveyor connection
- Support for open hood
- Support table with scratch-resistant bilaminated wood top
- 4 wheels kit for floor support
- Metal chest of 3 drawers with wheels
- EXACARB module - Activated carbon filtration system
|
Work area dimensions mm LxWxH
|
1130x550x570
|
|
Overall dimensions
mm LxWxH
|
1320x785x1460
|
|
Overall dimensions with support mm LxWxH
|
1320x785x2260 |
|
Average speed vertical laminar flow (m / s)
|
0.40 |
|
Average speed front barrier flow (m / s)
|
0.45
|
|
Total / expelled air volume (mc / h.)
|
1350/400
|
|
Weight in Kg
|
230 |
|
Absorbed power W
(230V 50Hz)
|
700 + 440
|
|
Protection fuses
|
2 x 5 AF (5 x 20) mm
|
| Watt / Lux lighting lamp |
1x36W 900 Lux |
| Network connection socket |
10A |
| UV germicidal lamp - accessory |
1 x 30W |
External dimensions of the product
| Height: |
1460 |
mm |
| Width: |
1320 |
mm |
| Depth: |
785 |
mm |
| Weight: |
230 |
Kg |