Logo Enrico Bruno srl
The items you find are only part of our catalog. We are in the process of completing the insertion

Code T6745055

Vertical laboratory autoclave mod. Lab 55


Vertical floor-standing autoclave with automatic and programmable sterilization cycle.

The external structure is made of 304 stainless steel, while the sterilization chamber and the lid are made of AISI 316 stainless steel, internally mechanically mirror polished.

The useful capacity of the sterilization chamber is 55 liters.

The exhaust vapors, after having been sterilized (optional), are condensed and conveyed into a special container, eliminating leaks and bad smells in the work environment.
The sterilization cycle is microprocessor controlled and fully programmable: the sterilization temperature and time can be defined by the user between 101 ° C and 138 ° C and from 1 to 99 minutes respectively.
The control system allows to memorize up to 100 different sterilization cycles.

The autoclave has a built-in printer-plotter in order to easily check and validate the cycle performed.



General features


Kind

Vertical floor-standing autoclave with automatic sterilization cycle

Maximum capacity

About 55 L

External structure

AISI 304 stainless steel

Work plan

AISI 304 stainless steel

Sterilization chamber

Cylindrical, in AISI 316L stainless steel, with mechanical cleaning and mirror polishing - 0.7 micron finish

Cover

AISI 316 L stainless steel with mechanical cleaning and mirror polishing, 0.7 micron finish, with central closure and equipped with a safety device against accidental opening, with double-lip gasket with active safety

Useful dimensions

Internal dimensions: 320 x 695 mm (Ø xh)

Plotter printer

Included, to print the text and the graph of the sterilization cycle and alarm states

Keyboard

Touch type, prot. IP66, waterproof

Display

Digital with continuous indication of the status and parameters of the sterilization cycle

Temperature probes

Thermal sensor for checking the heating resistance. The autoclave is designed for the installation of an additional PT100 probe for controlling the sterilization cycle.

Vapors and exhaust air

Collected in a waste tank, avoiding the loss of liquids and bad smells. The exhaust vapor can be filtered by a bacterial filter in order to preserve the safety conditions of the external environment

Water discharge

A tap allows the water present in the sterilization chamber to drain, in order to facilitate cleaning of the chamber.

Vacuum pump

Optional

Wheels

4 pivoting of which the two front ones are equipped with brakes


Sterilization cycle specifications

Sterilization cycle

Electronic control through microprocessor.

Programmable parameters

  • Sterilization temperature: from 101 ° C to 138 ° C
  • Sterilization time: from 1 to 99 minutes

Max. Working pressure

2.5 bar, at 138 ° C

Temperature accuracy

± 0.25 ° C to 121 ° C

Sterilization programs

No. 15, fully programmable

Heating system

Direct via heating resistor




Control systems and safety devices

Safety devices

  • Active safety: The lid is safe as long as the chamber is under pressure, even in the event of a black-out during the execution of the cycle.
  • Hydraulic safety valve (opening and closing of the drain) in case of overpressure
  • Maximum pressure switch

Control systems

 

  • Working temperature and sterilization time
  • Manometer to indicate the pressure with a range from 0 to 4 bar
  • Acoustic and visual alarm in case of malfunction
  • Check for the presence of sufficient water
  • Overpressure

 

Working conditions

Temperature

From + 15 ° C to + 40 ° C

Relative humidity

10 - 85% RH

 

Other data

Electrical feedind

230 V, 50/60 Hz; ± 10%

Power

3 KW

External dimensions

530 x 680 x 1110 mm (L x W x H)

Weight

85 Kg

Packaging dimensions

580 x 750 x 1320 mm (L x W x H) approximately

Weight with packaging

95 Kg

 

 

 

 

Optional accessories

T6055CT1

Perforated SS basket with dimensions 300 x 300 mm (dxh). (max cap n.2)

 

Additional self-locking lid.

T6055MAN

Quick cooling system at the end of the cycle

 

Vacuum pump application

T6055DRY

Automatic drying system at the end of the cycle

 

Debacterized breather system

 

Regulations

European directives

Directive 2004/108 / EC referring to EM compatibility

Directive 2006/95 / on the approximation of the laws of the Member States relating to electrical equipment intended for use within certain voltage limits

Machinery Directive 2006/42 / EC

Directive 97/23 / EC concerning pressure vessels (PED)

European regulations

EN61010-1 Safety requirements for electrical equipment for measurement, control and for use in the laboratory. General requirements.

EN61326-1 Electrical equipment for measurement, control and laboratory. Electromagnetic compatibility requirements. Part 1: general requirements. EMC requirements.

EN60204-1 Safety of machinery. Electrical equipment of machines P. 1: general rules



CONTACTS


ENRICO BRUNO s.r.l.
Via Duino 140
10127 Torino Italy
Tel. +39 011 3160808
Fax +39 011 3160753
PEC pec@pec.enrico-bruno.it
CF/P.IVA: 04891320014

WRITE TO US


For orders, prices requests and offers, technical service commerciale@enrico-bruno.it
For administration service amministrazione@enrico-bruno.it
For Direction – Purchase Department acquisti@enrico-bruno.it

FOLLOW US


Facebook Twitter Blogger