Vertical floor-standing autoclave with automatic and programmable sterilization cycle.
The external structure is made of 304 stainless steel, while the sterilization chamber and the lid are made of AISI 316 stainless steel, internally mechanically mirror polished.
The useful capacity of the sterilization chamber is 55 liters.
The exhaust vapors, after having been sterilized (optional), are condensed and conveyed into a special container, eliminating leaks and bad smells in the work environment.
The sterilization cycle is microprocessor controlled and fully programmable: the sterilization temperature and time can be defined by the user between 101 ° C and 138 ° C and from 1 to 99 minutes respectively.
The control system allows to memorize up to 100 different sterilization cycles.
The autoclave has a built-in printer-plotter in order to easily check and validate the cycle performed.
General features
Kind
|
Vertical floor-standing autoclave with automatic sterilization cycle
|
Maximum capacity
|
About 55 L
|
External structure
|
AISI 304 stainless steel
|
Work plan
|
AISI 304 stainless steel
|
Sterilization chamber
|
Cylindrical, in AISI 316L stainless steel, with mechanical cleaning and mirror polishing - 0.7 micron finish
|
Cover
|
AISI 316 L stainless steel with mechanical cleaning and mirror polishing, 0.7 micron finish, with central closure and equipped with a safety device against accidental opening, with double-lip gasket with active safety
|
Useful dimensions
|
Internal dimensions: 320 x 695 mm (Ø xh)
|
Plotter printer
|
Included, to print the text and the graph of the sterilization cycle and alarm states
|
Keyboard
|
Touch type, prot. IP66, waterproof
|
Display
|
Digital with continuous indication of the status and parameters of the sterilization cycle
|
Temperature probes
|
Thermal sensor for checking the heating resistance. The autoclave is designed for the installation of an additional PT100 probe for controlling the sterilization cycle.
|
Vapors and exhaust air
|
Collected in a waste tank, avoiding the loss of liquids and bad smells. The exhaust vapor can be filtered by a bacterial filter in order to preserve the safety conditions of the external environment
|
Water discharge
|
A tap allows the water present in the sterilization chamber to drain, in order to facilitate cleaning of the chamber.
|
Vacuum pump
|
Optional
|
Wheels
|
4 pivoting of which the two front ones are equipped with brakes
|
Sterilization cycle specifications
Sterilization cycle
|
Electronic control through microprocessor.
|
Programmable parameters
|
- Sterilization temperature: from 101 ° C to 138 ° C
- Sterilization time: from 1 to 99 minutes
|
Max. Working pressure
|
2.5 bar, at 138 ° C
|
Temperature accuracy
|
± 0.25 ° C to 121 ° C
|
Sterilization programs
|
No. 15, fully programmable
|
Heating system
|
Direct via heating resistor
|
Control systems and safety devices
Safety devices
|
- Active safety: The lid is safe as long as the chamber is under pressure, even in the event of a black-out during the execution of the cycle.
- Hydraulic safety valve (opening and closing of the drain) in case of overpressure
- Maximum pressure switch
|
Control systems
|
- Working temperature and sterilization time
- Manometer to indicate the pressure with a range from 0 to 4 bar
- Acoustic and visual alarm in case of malfunction
- Check for the presence of sufficient water
- Overpressure
|
Working conditions
Temperature
|
From + 15 ° C to + 40 ° C
|
Relative humidity
|
10 - 85% RH
|
Other data
Electrical feedind
|
230 V, 50/60 Hz; ± 10%
|
Power
|
3 KW
|
External dimensions
|
530 x 680 x 1110 mm (L x W x H)
|
Weight
|
85 Kg
|
Packaging dimensions
|
580 x 750 x 1320 mm (L x W x H) approximately
|
Weight with packaging
|
95 Kg
|
Optional accessories
T6055CT1
|
Perforated SS basket with dimensions 300 x 300 mm (dxh). (max cap n.2)
|
|
Additional self-locking lid.
|
T6055MAN
|
Quick cooling system at the end of the cycle
|
|
Vacuum pump application
|
T6055DRY
|
Automatic drying system at the end of the cycle
|
|
Debacterized breather system
|
Regulations
European directives
|
Directive 2004/108 / EC referring to EM compatibility
Directive 2006/95 / on the approximation of the laws of the Member States relating to electrical equipment intended for use within certain voltage limits
Machinery Directive 2006/42 / EC
Directive 97/23 / EC concerning pressure vessels (PED)
|
European regulations
|
EN61010-1 Safety requirements for electrical equipment for measurement, control and for use in the laboratory. General requirements.
EN61326-1 Electrical equipment for measurement, control and laboratory. Electromagnetic compatibility requirements. Part 1: general requirements. EMC requirements.
EN60204-1 Safety of machinery. Electrical equipment of machines P. 1: general rules
|